What a High-Level Disinfection Evaluation Revealed at One Urology Practice

Preventing Infection

What a High-Level Disinfection Evaluation Revealed at One Urology Practice

"Infection preventionists attributed several security breakdowns in infection control practices to a lack of education and understanding of high-level disinfection practices."

When infection preventionists (IPs) from Northwestern Medicine investigated the high-level disinfection (HLD) processes of a urology practice in July 2021, they identified several “breaches” in infection control practices.

Staff members responsible for HLD processing had been using expired chemical test strips and indicators during the reprocessing of reusable cystoscopes and prostate biopsy probes, according to a report published in the American Journal of Infection Control.

Infection preventionists attributed several security breakdowns in infection control practices to a lack of education and understanding of high-level disinfection practices.

The breaches included gaps in staff training, a lack of supplies, and inadequate staff supervision.

As part of their investigation, the IPs examined patient records from July 2019 through July 2021 to determine if any patients had a history of exposure to bloodborne pathogens, urinary tract infections or positive urine culture test results.

The report documented several cases of urinary tract infections following cystoscopy procedures linked to questionable HLD processes.

The IPs recommended that all cystoscopy procedures be suspended at the outpatient urology practice until the security breaches had been resolved.

The IPs also highlight the importance of ongoing leadership and infection prevention oversight in their report — HLD is a highly complex task requiring supervision to ensure patient and staff safety, they add.

A new, highly anticipated update to standards for reprocessing reusable endoscopic medical devices classifies flexible endoscopes — including cystoscopes and ureteroscopes — as "high-risk," necessitating sterilization rather than high-level disinfection to rid the scopes of contaminants that could harm patients.

The document, ANSI/AAMI ST91:2021, Flexible and semi-rigid endoscope processing in health care facilities, has been the subject of much debate in clinical and infection prevention circles since its release. It states that high-level disinfection (HLD) may not “reliably inactivate” certain types of microorganisms.

All flexible endoscopes should undergo sterilization to lower the risk of biofilm formation and enhance patient safety, according to the updated guidance from the Association for the Advancement of Medical Instrumentation (AAMI). Cleaning verification tests should be performed after each use of these scopes.

More Infection Prevention Articles
Why Medical Technologists Should be Considered for IP Roles
Prevention Strategies
The COVID-19 pandemic has exacerbated the healthcare worker shortage and infection preventionists are no exception.
What to Know About ANSI/AAMI ST108
Prevention Challenges
“Water is actually really complex in its chemistry and depending on where you are geographically, the water quality can be very different,” says the co-chair of a working group authoring the new standard.
More From Single-Use Endoscopy
Study Finds Single-Use Gastroscopes Effective in Advanced Procedures

Emerging Technologies

Single-use scopes are known to be capable of diagnostic procedures, but little research had tested their therapeutic capabilities.

Study: Patients Prefer Option of Selecting Endoscopes

Public Health

A study out of Malaysia showed healthcare consumers, if given the choice, preferred a single-use endoscope over a reusable one or a doctor’s preferred device.

See what made ERCI

Public Health

Successful device reprocessing requires clear instructions to keep patients safe, the organization says in its closely watched annual report.